Digital X-ray Image Acquisition
System
- Digital Detector Trixell Pixium®4600 - CsI:Tl/a-Si
- 43x43 cm (17x17 inch) active area
- 143 µm pixel pitch
- 3.5 cycles/mm detector Nyquist frequency
- DQE
>55% @ 1.0 cycles/mm
>45% @ 2.0 cycles/mm
>25% @ 3.0 cycles/mm
- 14-bit dynamic range
Operator Console
- Controls detector and X-ray generator
- Patient demographics via keyboard or HIS/RIS
- Anatomically Programmed Radiography (APR)
- Automatic Exposure Control (AEC)
- Automatic collimator remote control
- DICOM 3.0 compliant
- Processing algorithms (exam-specific)
- Print layout film editor
- 18-inch high resolution TFT LCD
Digital Radiography System
Ceiling mounted X-ray tube suspension
- Minimum focal spot/ceiling distance: 93 cm (36.6 inch)
- Focal spot/telescopic axis distance: 35 cm (13.8 inch)
- Exploitation area: 437x295 cm (172.0x116.1 inch)
- Vertical movement speed: 8 cm/s (3.1 inch/s)
- Tube vertical travel: 160 cm (63.0 inch)
- Tube rotation around telescopic axis: 360°
- Tube rotation around horizontal axis: +/- 115°
- Longitudinal, lateral and vertical brake: negative - 24 Vdc
- Tube rotation brake: negative -24 Vdc
- FFD, tube rotation angle data on display
- Automatic collimator with remote control
- Automatic source detector alignment
- Electrical power: 115/230 VAC, single phase
- Max power required: 700 VA
- Weight: 350 Kg (771.6 lbs)
- Typical room ceiling height: 280 cm (110 inch)
Wall-mounted digital bucky (vertical
stand)
- Bucky for digital detector
- Removable 12:1, 80 l/cm grid
- Electrical power: 120/240 VAC, single phase
- Frequency: 50/60 Hz
- Max power required: 500 VA
- Dimensions: H: 187,5 L: 64 W: 36 cm
- Vertical shift: 110 cm Weight: 102 Kg
- Dimensions (cm): H: 187.5; L: 64; W: 36
(inch): H: 73.8; L: 25.2; W: 14.2
- Weight: 102 Kg (224.9 lbs)
X-ray Tube
- 32/72 kW (1)
- Exposure voltage: 40÷150 kVp
- Maximum anode heat content: 350 kHU
- Focal spots: 0,6-1,0 mm
- Anode rotation speed: 9000 rpm
X-ray Generator
- 65 kW high frequency (100 kHz) output (2)
- 40-150 kVp
- Dual speed starter
- mAs range: 0.5-1,000 mAs
- Digital interface
- 400 VAC, 480 VAC, 3 phase
Optional Components
- Automatic Operator Log-In
- Laser Printer Interface
- X-ray generator control console
- Patient stand stretcher and grip
- Digital Keypad
- BAR code reader
- CD-R
|