Digital X-ray Image Acquisition
System
- Digital Detector Trixell Pixium®4600 - CsI:Tl/a-Si
- 43x43 cm (17x17 inch) active area
- 143 µm pixel pitch
- 3.5 cycles/mm detector Nyquist frequency
- DQE
>55% @ 1.0 cycles/mm
>45% @ 2.0 cycles/mm
>25% @ 3.0 cycles/mm
- 14-bit dynamic range
Operator Console
- Controls detector and X-ray generator
- Patient demographics via keyboard or HIS/RIS
- Anatomically Programmed Radiography (APR)
- Automatic Exposure Control (AEC)
- Automatic collimator remote control
- DICOM 3.0 compliant
- Processing algorithms (exam-specific)
- Print layout film editor
- 18-inch high resolution TFT LCD
Wall-mounted U-arm
- U-arm linked X-ray tube and FPD
- Focal distance to Bucky: 100-180 cm (40-72 inch)
- Horizontal beam height to floor: 50-160 cm (20-63 inch)
- Swivel arm rotation: +/- 135°
- Motorized SID
- Bucky tilt angle: +/- 45°
- Electromagnetic positioning brakes
- Removable 12:1, 80 l/cm grid
- Optional patient stretcher and grip
X-ray Tube
- 32/72 kW (1)
- Exposure voltage: 40÷150 kVp
- Maximum anode heat content: 350 kHU
- Focal spots: 0,6-1,0 mm
- Anode rotation speed: 9000 rpm
X-ray Generator
- 65 kW high frequency (100 kHz) output (2)
- 40-150 kVp
- Dual speed starter
- mAs range: 0.5-1,000 mAs
- Digital interface
- 400 VAC, 480 VAC, 3 phase
Optional Components
- Automatic Operator Log-In
- Laser Printer Interface
- X-ray generator control console
- Patient stand stretcher and grip
- Digital Keypad
- BAR code reader
- CD-R
|